Thermal Measuring Tube TA Di - standard - with integrated transducer

Thermal Measuring Tube TA Di -
standard - with integrated transducer

Thermal Measuring Tube TA Di - 
standard - with integrated transducer

Thermal Measuring Tube TA Di - standard - with integrated transducer

Measuring tube TA Di with inside diameter 8, 16, 21.6, 27.2, 35.9 41.8 and 53.1 mm and transducer U10b in sensor connection housing. Via web interface, measuring instruments having a WiFi interface can be accessed by a web-enabled device, performance data can be checked and transferred and the device settings can be customised according to the measuring task. The unit can be used for a variety of gases even in ex-applications categories 3G and 3D (zone 2 and zone 22) That way there is no need for recalibration or return to the manufacturer when there’s a change of gas. The corresponding gas characteristics stored in the device simply have to be selected on the sensor.

Probe material: stainless steel
Working temperature range -10 ... +140 °C
Pressure resistance up to 16 bar

Applications:

Fixed measuring of compressed air and gas consumption of oxygen, nitrogen, argon e.g. in welding technologies, leakage flow, outgoing air, burner supply air, monitoring inerting of core processes, in air in low vacuum range at pressures greater than 200 hPa absolute.

Output signals:

  • 2 analog outputs 4-20 mA - parameters freely adjustable
  • relay for quantity pulse
  • WiFi interface (optional)
  • M-Bus (optional). Click here for more M-Bus information

Configurable via an optional keypad in the housing lid or via WiFi interface.

 

Examples for common applications:

Thermal post-combustion after chemical installations
Energy management DIN EN ISO 50001/ measurement of compressed air consumption
Sludge aeration air in sewage plants
Leakage measurement at large ball valves in gas pipelines

Measuring range

0.6 ... 23967 l/min or 0.04 ... 1196 m³/h

Pressure resistance

up to 16 bar / 1.6 MPa above atmospheric

Media temperature

-10 ... +140 °C

Media

air (compressed air), nitrogen, methane, natural gas, propane, butane
argon, carbon dioxide, sulphur hexafluoride, helium, hydrogen, landfill gas, oxygen, …

Displayed indicators

Standard flow velocity [Sm/s],
Standard flow rate [Sm³/h] or [Sl/min],
Mass flow rate [kg/h],
Standard volume [Sm³],
Temperature [° C, ° F].
With optional M-Bus: standard flow rate, mass flow rate, standard volume.

IP protection type

IP65 (connection housing)

Type of protection

ATEX category 3G and 3D (Zone 2 and 22) (optional)

Sensor output

2 analog outputs 4 - 20 mA , linear, freely configurable
relay for quantity pulse
WiFi (optional)
M-Bus (optional)